Waters Alliance e2695 Operator's Manual page 204

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

7 Maintenance
Vial of needle-wash solvent being used or alternatively 100% alcohol
(methanol or isopropanol)
Replacement one-piece needle-wash frit
Immerse the replacement frit for several minutes in the vial of the
Tip:
needle-wash solvent being used, or alternatively in alcohol before beginning
the replacement procedure.
To replace the lower frit:
1.
Open the sample carousel door, as shown below in the figure.
2.
Remove all carousels that restrict access to the lower retainer.
Use the Door is Open dialog box to move the carousel carrier to the
Tip:
"A" position. See
3.
Power-off the separations module, and disconnect it from the electrical
outlet.
4.
Find the lower frit retainer, which is inside the sample compartment on
the top right side, approximately 4 inches within the cavity.
Locate the lower frit retainer:
Internal sample compartment cavity
(viewed from front left-side of door)
Lower
frit
retainer
204
July 29, 2013, 715003794 Rev. B
"Load carousels" on page 111
for details.