3 Prepare the e2695 Separations Module for Operation
Preparing the sample management system:
Start
Purge the sample
management
system
Prime the needle
wash pump

Purge the system

Caution:
intermediate solvent such as water when you change from buffers to
high-organic-content solvents.
Purge the sample management system on these occasions:
You primed the solvent management system.
You changed solvents.
You see bubbles in the syringe.
You start using the separations module at the beginning of each day.
To purge the sample management system:
1.
In the Main screen, press Menu/Status.
2.
In the Status screen's Composition fields, specify the solvent
composition.
3.
Press Direct Functions to display the Direct Functions menu.
4.
From the Direct functions menu, select Purge Injector, and then press
Enter. The Purge Injector dialog box appears.
108
July 29, 2013, 715003794 Rev. B
Adjust the seal pack
Load the carousels
Separations
module is ready
to operate
To avoid precipitating salts in the separations module, use an