2 Install the e2695 Separations Module
Screwdriver, T-20 TORX
Wrench, 7 mm
To install the drip tray on the detector (applies to 2424, 2489, and 2998 only):
1.
Disconnect all cabling and other connections from the detector.
2.
Open the front door of the detector.
3.
Rest the detector on its rear panel, so that the front panel is facing up.
4.
If the detector is a 2424, remove the barbed fitting from the underside of
the detector, as this interferes with the mounting of the drip tray.
5.
Remove the T-10 screw located on the bottom right corner of the detector
front panel and set it aside.
A and C series drip tray mounting location on detector:
Drip tray
mounting flange
6.
Position the drip tray so that the mounting hole for the T-10 screw you
removed is aligned between the mounting ears on the right end of the
drip tray.
7.
Push the mounting slot on the opposite side of the drip tray into the
mounting flange on the detector until the drip tray locks into place.
8.
Insert the T-10 screw you removed earlier in the hole between the drip
tray mounting ears and secure the drip tray to the detector.
54
July 29, 2013, 715003794 Rev. B
T-10 screw