Connect The Column - Waters Alliance e2695 Operator's Manual

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

2 Install the e2695 Separations Module
Needle wash and plunger seal wash line color code:
Wash line
Needle-Wash Inlet
Needle-Wash Outlet
Sample-Loop Waste
Plunger Seal-Wash
Inlet
Plunger Seal-Wash
Waste
To install the needle-wash and plunger seal-wash reservoirs:
1.
Place the needle-wash reservoir and plunger seal-wash reservoir in the
drip tray atop the separations module.
2.
Place the needle-wash inlet line (green) in the needle wash reservoir.
3.
Place the needle-wash outlet line (yellow or clear) and the sample loop
waste line (clear) in a suitable waste container.
4.
Remove the diffuser from the plunger seal-wash inlet line (clear, labeled
Pump Wash In), insert the line through a reservoir cap, reinstall the
diffuser on the end of the line, and install the cap on the plunger
seal-wash reservoir.
5.
Place the plunger seal-wash waste line (clear) in a suitable waste
container.

Connect the column

The red outlet tubing of the sample management system, which connects to
the column inlet, is located behind the right-hand side panel.
When you are connecting the separations module to a column switching valve,
connect the red outlet tubing to the column switching valve inlet. When you
are connecting the separations module to an external autosampler and
configuring for operating in Operate Gradient by Event In mode, connect the
red outlet tubing to the inlet of the autosampler.
62
July 29, 2013, 715003794 Rev. B
Color code
Green
Yellow or clear
Clear
Clear (labeled "Pump Wash
In")
Clear
Inside diameter
(inches)
1/16
1/16
1/16
1/8
1/8