Set I/O Parameter Values - Waters Alliance e2695 Operator's Manual

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

6 Create Methods, Sample Sets, and Sample Templates
Column parameters: (Continued)
Parameter
Column
Temperature
Range
Column
Selection
3-Column
Valve
Column
Information
Column Info
(screen key)
Equilibration

Set I/O parameter values

You set the I/O parameter values when you want to use analog signals to
notify other instruments of the status of the separations module. You can do
any of these things:
Notify a detector that an injection began.
Turn on a heating plate or a stirring bar in a solvent reservoir.
Notify a third-party data system or integrator that an event occurred.
Activate a switching valve.
Sound a buzzer.
162
July 29, 2013, 715003794 Rev. B
Function
Sets the maximum allowable
variation in column temperature. If
the temperature variation exceeds
the range, the alarm condition
selected in the adjacent box is
triggered.
Selects a column position if a
column selection valve is installed.
Select No Change when you
Tip:
link methods in a sample set and
do not want to change the column
from the previous method.
Displays the column information
you enter using the Column Info
screen key.
Allows you to enter or modify the
column information field.
Allows you to specify an
equilibration time that will be
applied whenever the application
of the separation method involves a
change in the column switch
position.
Value range
±10 °C in 1 °C
increments
Position 1 through
Position 3, or No
Change
N/A
30 characters
< 999.99 minutes