Add Hardware Options; Install An Optional Sample Loop - Waters Alliance e2695 Operator's Manual

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

2 Install the e2695 Separations Module
As with IEEE-488 control, triggering of Waters Ethernet-controlled devices,
such as 2489, 2998, 2414, and 2424 detectors, occurs over the Ethernet cable,
and so requires no external I/O cables.
Waters 2475 Fluorescence Detector, v2.01 requires an external I/O
Exception:
trigger cable under Ethernet control.
Configure the Ethernet port
On the Configuration screen, select Auto/Ethernet for the separations module.
See
"Configure the operating mode" on page 98
configuring the Ethernet port.

Add hardware options

Depending on your applications, you can install the following options on the
separations module:
Larger sample loop
Column heater
Column heater/cooler
Larger or smaller syringe
Column selection valve

Install an optional sample loop

To inject sample volumes greater than 100 µL, install an optional sample loop.
To do so, you must perform these tasks:
Remove the right-hand side panel
Install the sample loop
Refit the right-hand side panel
Required materials
Screwdriver, #2 Phillips
Wrenches, 5/16-inch, two
Wrench, adjustable
Sample Loop Option Kit
78
July 29, 2013, 715003794 Rev. B