Prime The Plunger Seal-Wash Pump - Waters Alliance e2695 Operator's Manual

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

Prime the plunger seal-wash pump

The prime plunger seal-wash test automatically primes the solvent
management system plunger seal-wash pump. Perform this test if you discern
a lack of flow in the plunger seal-wash system or if you are changing the
seal-wash solvent. If required, use an intermediate solvent to prevent
miscibility or precipitation problems when changing solvents.
Required materials
Seal-wash solution
Tubing adapter (startup kit)
Syringe (use only if no solvent is in the seal-wash pump)
To prime the plunger seal wash:
1.
Ensure that the plunger seal-wash supply line (labeled "Inlet") is in the
seal-wash solvent bottle and that the plunger seal-wash waste line
(clear) is in a suitable waste container.
2.
Attach the tubing adapter (startup kit) to the syringe.
3.
Fill the syringe with the seal-wash solution, and then connect the
syringe assembly to the seal-wash inlet tube.
4.
Press the Prime SealWsh screen key in the Diagnostics screen.
5.
In the Prime Seal Wash dialog box, press Start to begin the seal-wash
procedure.
6.
Push on the syringe plunger to push seal-wash solvent through the
system.
7.
When seal-wash solvent flows out the seal-wash waste tube, press the
Halt screen key.
Result:
the test is not halted.
8.
Press Close to return to the Diagnostics screen.
For more information about priming the plunger seal-wash pump, see
the plunger seal-wash pump" on page
The seal-wash pump prime will time-out after fifteen minutes if
Perform main diagnostic tests
106.
July 29, 2013, 715003794 Rev. B
"Prime
223