Bus Communication; Performance - Siemens SITRANS FC430 Operating Instructions Manual

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

13.4

Bus communication

Table 13- 5
Description
Manufacturer ID
Device ID
HART protocol revision
Device revision
Number of device variables
Physical layers supported
Loop-powered
SIMATIC PDM SW
13.5

Performance

Table 13- 6
Description
Process media
Process media temperature
Ambient temperature
Process media pressure
Process media density
Reference device orientation
Table 13- 7
Description
Sensor size
Qmin - minimum flowrate [kg/h] (lb/m)
Qnom - nominal flowrate [kg/h] (lb/m)
Qmax - maximum flowrate [kg/h] (lb/m)
Max. zero point stability [kg/h]
Measuring accuracy [%]
Repeatability error [%]
FC430 with HART
Operating Instructions, 05/2015, A5E03361511-AF
HART communication
Specification
42 (2A Hex)
34 (22 Hex)
7.2
1
12
FSK
No
6.0 SP2 and higher
Reference conditions
Massflow accuracy
13.4 Bus communication
More information
Manufacturer ID parameter
Device type parameter
HART protocol revision parameter
Device revision parameter
Number of process values, both measured
and derived
Frequency Shift Keyed
4-wire device
Software version
Specification
Water
20 °C (68 °F)
25 °C (77 °F)
2 bar (29 psi)
0.997 g/cm
(62.2 lb/ft
)
3
3
Horizontal installation, tubes down, flow in direc-
tion of arrow on casing, see Installing/Mounting
(Page 48).
Specification
DN 15
DN 25
20
200
(0.735)
(7.35)
3700
11 500
(136)
(422.6)
6 400
17 700
(14 110)
(39 022)
±0.2
±2.0
±0.10
±0.05
Technical data
DN 50
DN 80
750
900
(27.6)
(33.1)
52 000
136 000
(1 911)
(4 997)
70 700
181 000
(155 867)
(399 036)
±7.5
±18.0
195

Advertisement

Table of Contents
loading