Troubleshooting/Faqs; Diagnosing With Pdm; Troubleshooting Sensor-Related Problems - Siemens SITRANS FC430 Operating Instructions Manual

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

Troubleshooting/FAQs

12.1

Diagnosing with PDM

SIMATIC PDM is a suitable tool for diagnosing the device.
You can use SIMATIC PDM to read all available parameters to a table for analyzing offline,
view online/actual process values and online/actual diagnostic information.
Requirements
The following procedure must be completed before diagnosing:
● Installation of PDM and PDM device driver
● Connection of HART interface
Refer to "Commissioning with PDM" (Page 88).
Diagnosing with PDM
Online process values are available under menu "View->Process Values".
Online diagnostic information is available under menu "View->Device Status"
12.2

Troubleshooting sensor-related problems

Incorrect and unstable measurements, especially at low flows, are typically a result of an
unstable zero point due to:
● Incorrect installation
● Bubbles in the liquid
● Vibrations/Cross talk
● Solid particles settling in the liquid
In the following a 4-step guide to troubleshooting is provided:
Step 1
Step 2
Step 3
Step 4
The guide will enable you to trace the reason for incorrect measurements and to improve the
application.
FC430 with HART
Operating Instructions, 05/2015, A5E03361511-AF
Preliminary application inspection
Zero point adjustment
Measurement error calculation
Application improvement
12