Certificates and approvals HART

Table 13- 39 Certificates and approvals
Description
ATEX
FC430 with HART
Operating Instructions, 05/2015, A5E03361511-AF
13.15 Certificates and approvals HART
Specification
FCT030 transmitter (can be installed in Zone 1
for gas and Zone 21 for dust):
Certificate SIRA 11ATEX1342X:
II 2(1) GD
Ex d e [ia Ga] IIC T6 Gb Ta = -40°C to +60°C
Ex tb [ia Da] IIIC T85°C Db
FCS400 sensor + DSL (can be installed in Zone
1 for gas and Zone 20/21 for dust):
ATEX Certificate: SIRA 11ATEX1341X
II 1/2 G
1D
2D
For gas:
Ex d ia IIC T* Ga/Gb
Ex d IIC T* Ga/Gb
For dust:
Ex ta IIIC T* °C Da
Ex tb IIIC T* °C Db
Ta = -40°C to +60°C
* Temperature class (dependent on the process
temperature and the ambient temperature, see
Special Conditions for Safe Use, Section 2.3)
FC430 compact system (can be installed in Zone
1 for gas and Zone 21 for dust):
ATEX certificate SIRA 12ATEX1102X
II 1/2 (1) G
II 2 D
Ex d e ia [ia GA] IIC T* Ga/Gb Ta = -40°C to ** °C
Ex tb [ia Da] IIIC T**°C Db
* Temperature class (dependent on the "Maxi-
mum Process Temperature")
** Upper ambient temperature (dependent on the
"Maximum Process Temperature")