Siemens SITRANS FC430 Operating Instructions Manual page 175

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

ID
Diagnostic
38
Temperature measurement fault Turn off the power, wait 5 seconds and turn on
39
40
41
46
Invalid calibration data
47
Invalid compensation data
49
Malfunction in Pickup Amplitude Contact Siemens customer support
50
51
55
Malfunction in sensor driver
56
57
58
Unstable driver oscillation
71
Parameter storage malfunction
72
Internal error in sensor
73
74
75
76
77
150
Sensor signal disrupted
157
Safety alarms
158
HART cable break
159
Internal error in transmitter
171
Product FW incompatible
172
Transm. FW incompatible
173
Sensor FW incompatible
174
HMI FW incompatible
197
Current output cable break
203
Current output cable break
209
Current output cable break
213
Invalid dosing configuration
FC430 with HART
Operating Instructions, 05/2015, A5E03361511-AF
Action
the power again. If the failure continues, then
contact Siemens customer support
Contact Siemens customer support for recali-
bration
Contact Siemens customer support
Contact Siemens customer support
Contact Siemens customer support
Turn off the power, wait 5 seconds and turn on
the power again. If the failure continues then
contact Siemens customer support
Contact Siemens customer support
Turn off the power. Unplug and reconnect the
sensor cable. Restore power. If the failure
continues, then contact Siemens customer
support
The system has detected a safety-related
alarm in Safe Operation. The device is in safe-
ty alarm state. All alarms must be cleared to
bring the unit to Safe Operation mode.
Check channel 1 current output cable connec-
tion
Turn off the power, wait 5 seconds and turn on
the power again. If the failure continues, con-
tact Siemens customer support
Firmware update failed
Update rejected - transmitter FW incompatible
Update rejected - sensor FW incompatible
Update rejected - HMI FW incompatible
Check channel 2 current output cable connection
Check channel 3 current output cable connection
Check channel 4 current output cable connection
The Two Stage Dosing controls two valves with
use of two Signal Outputs. To ensure valid
configuration either Stage 1 Primary Open or
Stage 2 Secondary Open must be set to 0 and
either Stage 1 Primary Close or Stage 2
Secondary Close must be set to Amount
Alarms and system messages
10.2 Alarm messages
Effect on output
Yes
Yes*
Yes*
Yes*
Yes*
Yes*
Yes
173