Alarms and system messages
10.2 Alarm messages
ID
Diagnostic
140
Totalizer 2 above alarm limit
143
Totalizer 2 below alarm limit
144
Totalizer 3 above alarm limit
147
Totalizer 3 below alarm limit
148
Transm. temp. above alarm limit Check process conditions or align limit to nor-
149
Transm. temp. below alarm limit
153
Current output below Scaling
154
Current output above Scaling
192
Dosing time overrun
193
Dosing quantity overrun
194
Invalid proc. val. during dosing
195
Current output below Scaling
196
Current output above Scaling
198
Frequency output below Scaling
199
Frequency output above Scaling Check process conditions or align limit to nor-
200
Pulse overflow
201
Current output below Scaling
176
Action
Check process conditions or align limit to nor-
mal operation. Adjust parameter "Upper Limit
Alarm"
Check process conditions or align limit to nor-
mal operation. Adjust parameter "Lower Limit
Alarm"
Check process conditions or align limit to nor-
mal operation. Adjust parameter "Upper Limit
Alarm"
Check process conditions or align limit to nor-
mal operation. Adjust parameter "Lower Limit
Alarm"
mal operation. Adjust parameter "Upper Limit
Alarm"
Check process conditions or align limit to nor-
mal operation. Adjust parameter "Lower Limit
Alarm"
Check process conditions or align limit to nor-
mal operation. Adjust channel 1 parameter
"Lower Scaling"
Check process conditions or align limit to nor-
mal operation. Adjust channel 1 parameter
"Upper Scaling"
Check installation. If ok, increase "Duration
Time"
Check installation. If ok, decrease "Overrun
Value"
Check installation for abnormal operating
conditions.
If the failure continues for several dosings,
contact Siemens customer support
Check process conditions or align limit to nor-
mal operation. Adjust channel 2 parameter
"Lower Scaling"
Check process conditions or align limit to nor-
mal operation. Adjust channel 2 parameter
"Upper Scaling"
Check process conditions or align limit to nor-
mal operation. Adjust channel 2 parameter
"Flow Value Low"
mal operation. Adjust channel 2 parameter
""Flow Value High"
Pulse output insufficient pulse separation.
Increase "Amount Per Pulse" or reduce "Pulse
Width" on channel 2
Check process conditions or align limit to nor-
mal operation. Adjust channel 3 parameter
"Lower Scaling"
Effect on output
Operating Instructions, 05/2015, A5E03361511-AF
FC430 with HART