Step 5: Finishing The Transmitter Connection - Siemens SITRANS FC430 Operating Instructions Manual

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

Connecting

5.11 Step 5: Finishing the transmitter connection

Load [Ω]
5000
10000
20000
50000
100000
Note
Load
Signal output: < 500 Ω at 14 to 24 VDC (active), 14 to 30 VDC (passive)
Relay output: 30 VAC/VDC, 100 mA
Passive signal input: 15 to 30 VDC, 2 to 15 mA
5.11
Step 5: Finishing the transmitter connection
Connection check-up
1. Check individual wire installation by tugging firmly.
2. Firmly tighten cable glands and insert blanking plugs in unused cable entries.
3. Remove O-ring from lid.
4. Reinstate lid and screw in until mechanical stop. Wind back lid by one turn.
5. Mount O-ring by pulling it over the lid and tighten lid cover until you feel friction from the
O-ring on both sides. Wind lid further by one quarter of a turn to seal on the O-ring.
6. Reinstate and tighten lid lock screw.
7. Ensure that moisture does not penetrate to inside of electronics housing by creating a
drip loop (bend cables downward) immediately before cable glands.
Loss of degree of protection
Damage to device if the enclosure is open or not properly closed. The degree of protection
specified on the nameplate or in Chapter "Design (Page 201)" is no longer guaranteed.
• Make sure that the device is securely closed.
72
CAUTION
Voltage (active) [V]
20.42
21.13
21.51
21.74
21.82
Operating Instructions, 05/2015, A5E03361511-AF
Voltage (passive) [V]
21.80
21.85
21.88
21.89
21.90
FC430 with HART