Operating
7.1 Local display (HMI)
Changing the resolution:
1. Select decimal point by pressing
2. Move decimal point by pressing
(moves decimal point to the right).
In order to change the resolution of the process value shown in the operation view (for
example massflow), change the resolution of one configuration parameter for this process
value (for example "Low Flow Cut-off" (menu item 2.2.1.2)). Any changes in resolution will
change the resolution of all configuration parameters for this process value as well.
The resolution can also be changed by setting the decimal places parameter for the selected
process value, for example the decimal places for process value Massflow is defined in
parameter menu item 2.2.1.8.
Numeric parameter read only view
Numeric parameters in read only view are displayed as shown here.
Figure 7-7
The read only view is shown if you don't have access to edit parameters. The view shows
the set value. Press
128
Escape
The set value
Parameter name
Parameter item number
Help text describing the parameter function.
Numeric parameter read only view
to escape the view.
and
keys.
key (moves decimal point to the left) or
Operating Instructions, 05/2015, A5E03361511-AF
key
FC430 with HART