4
General Maintenance
To Connect the Micro-Ion Vacuum Gauge
The Micro-Ion vacuum gauge is optional for EI operation. It is
operating the MSD in CI mode.
Materials needed
• Micro-Ion vacuum gauge (G3397A)
• Power cord
• Micro-Ion vacuum gauge cable (G3170-60805)
• CI flow module cable-Smart Card to CI flow (G3170-60808)
Procedure
1
2
3
92
Use the provided KF-16 clamp to install the Micro-Ion vacuum gauge to the
rear of the analyzer chamber.
Connect the gauge cables as shown in
Pump down the MSD.
Figure
5975 Series MSD Troubleshooting and Maintenance Manual
required
5.
for