To Connect The Micro-Ion Vacuum Gauge - Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual

5975 series
Table of Contents

Advertisement

4
General Maintenance

To Connect the Micro-Ion Vacuum Gauge

The Micro-Ion vacuum gauge is optional for EI operation. It is
operating the MSD in CI mode.
Materials needed
• Micro-Ion vacuum gauge (G3397A)
• Power cord
• Micro-Ion vacuum gauge cable (G3170-60805)
• CI flow module cable-Smart Card to CI flow (G3170-60808)
Procedure
1
2
3
92
Use the provided KF-16 clamp to install the Micro-Ion vacuum gauge to the
rear of the analyzer chamber.
Connect the gauge cables as shown in
Pump down the MSD.
Figure
5975 Series MSD Troubleshooting and Maintenance Manual