Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual page 149

5975 series
Table of Contents

Advertisement

To reinstall the heater and sensor in the mass filter
Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Mass filter heater assembly (G1099-60172)
Procedure
1
2
3
4
5
Do not touch the mass filter contact leads. This could cause ESD damage to the side
CA UT IO N
board.
5975 Series MSD Troubleshooting and Maintenance Manual
Large (8650-0030)
Small (8650-0029)
Unpack the new mass filter heater assembly
temperature sensor, and heater block are already assembled.
Install the heater assembly on top of the mass filter radiator.
Connect the heater and temperature sensor wires to the feedthrough
board.
Close the analyzer chamber.
Pump down the MSD.
General Maintenance
(Figure
25). The heater,
4
149