To Refill The Foreline Pump - Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual

5975 series
Table of Contents

Advertisement

4
General Maintenance

To Refill the Foreline Pump

Standard foreline pump only
Materials needed
• Foreline pump oil (6040-0621) – approximately 0.28 L required
• Funnel (9301-6461)
• Gloves, oil- and solvent-resistant
• Screwdriver, flat-blade, large (8730-0002)
• Drain plug O-ring (if required) (0905-1515)
• Hex key, 5-mm (8710-1838)
Procedure
The foreline pump can cause burns if touched when operating. It has a safety shield
WARNING
to prevent the user from touching it.
1
2
3
4
5
6
7
8
100
Drain the foreline pump. See
Reinstall the drain plug. If the old O-ring appears worn or damaged,
replace it.
Remove the propping object from under the pump motor.
Add foreline pump oil until the oil level in the window is near, but not
above, the upper line. The foreline pump requires approximately 0.28 L of
oil.
Wait a few minutes for the oil to settle. If the oil level drops, add oil to bring
the oil level near the upper line.
Reinstall the fill cap.
If necessary, slide the foreline pump back under the analyzer chamber.
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
Pump down the MSD.
page
98.
5975 Series MSD Troubleshooting and Maintenance Manual

Hide quick links: