7
Analyzer
Filaments
Two filaments
EI ion source. The
The emission current heats the filament causing it to emit electrons which
ionize the sample molecules. In addition,
DC bias voltage. The bias voltage determines the energy on the electrons,
usually –70 eV.
The CI ion source has only one filament of a different design from the EI
filaments. A "dummy" filament provides connections for the Filament 2 wires.
The filament is shut off automatically if there is a general instrument
shutdown. Three parameters affect the filaments: filament selection
(Filament), filament emission (Emission) current, and electron energy
(EIEnrgy).
Filament selection
The filament selection parameter (Filament) selects which filament in the ion
source is active. In the CI source, it is always Filament 1.
Sometimes, one EI filament will give better performance than the other. To
select the better of the two filaments, run two autotunes, one with each
filament. Use the filament that gives the best results.
Emission current
The filament emission current (Emission) is variable between 0 and –315 µA,
but should be set to the software default for normal operation.
Electron energy
The electron energy (EIEnrgy) is the amount of energy on the ionizing
electrons. It is determined by the bias voltage; –70 VDC bias on the filament
causes emitted electrons to possess –70 eV (electron volts). This value is
adjustable from –5 to –241 VDC, but for normal operation, set this parameter
to 70.
204
(Figure
46) are located on opposite sides of the outside of the
active
filament carries an adjustable AC emission current.
5975 Series MSD Troubleshooting and Maintenance Manual
both
filaments have an adjustable