Install The Micro Vacuum Degasser (G1379A) - Agilent Technologies 1100 Series System Manual

Capillary lc system
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Connect the power cable to power only after you have finished the hardware
NOTE
Carefully route all capillaries so that they are not crushed or broken by module front covers.
Avoid excessive bending. Refer to
6 Connect the 1/8 inch plastic EMPV waste tube to the barbed waste fitting of

Install the Micro Vacuum Degasser (G1379A)

WARNING
installation of all modules in the stack.
1 Ensure the line power switch at the front of the micro vacuum degasser is off.
2 Place the degasser on top of the pump. Make sure that the two modules are
3 Connect one end of the remote cable (5061-3378) to the rear of the degasser.
4 The degasser accessory kit has a set of 4 solvent tubes (G1322-67300). Each tube is
1100 Series Capillary LC System Manual
the EMPV. Route the waste tube to an appropriate waste container.
interlocked correctly.
Connect the other end of the cable to the remote port at the rear of the pump.
labeled A, B, C or D. Connect each solvent tube to its intended OUTLET channel
port on the degasser.
Installing your Capillary LC System 1
Chapter 2
for advice on handling capillaries.