To Refill The Ci Calibrant Vial - Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual

5975 series
Table of Contents

Advertisement

To Refill the CI Calibrant Vial

Materials needed
• PFDTD calibrant (8500-8510)
Procedure
1
2
3
4
5
6
Do not rinse the vial with any solvents. Never expose the inside of the vial to
CA UT IO N
chlorinated solvents or isopropyl alcohol or water — this will result in severe loss of CI
sensitivity.
7
8
9
10 Reinstall the capillary column.
11 Pump down the MSD.
12 Purge the calibration valve. See
After removing the calibrant vial, you must purge the calibration valve. Failure to do so
CA UT IO N
will result in severe contamination of the ion source and damage to the filament and
electron multiplier.
5975 Series MSD Troubleshooting and Maintenance Manual
Set the reagent gas flow to Gas Off.
Vent the MSD.
Remove the capillary column from the GC/MSD interface.
Pull the MSD away from the GC to expose the calibration vial and valve.
See
page
103.
Loosen the collar holding the calibration vial in place. Do not remove the
collar.
Remove the calibrant vial. See
Fill the vial no higher than the bottom of the internal tube with fresh
PFDTD calibrant (8500-8510).
Replace the vial and tighten the collar.
Reposition the MSD next to the GC. See
Figure
33.
page
105.
page
111.
5
CI Maintenance
169