4
General Maintenance
To Reconnect the MSD to the 6850 GC
Materials needed
• Wrench, open-end, 1/4-inch 5/16-inch (8710-0510)
• Screwdriver, Torx T-15 (8710-1622)
Procedure
1
2
Figure 10
106
Position the MSD so that it is near the GC.
If not installed, install the GC/MSD interstitial panel (part number
G2589-00031) onto the midsupport and front-support legs of the right side
of the MSD. Use three M4 T-15 screws. See
Mounting holes
GC/MSD interstitial panel mounting holes (turbo pump shown)
5975 Series MSD Troubleshooting and Maintenance Manual
Figure