Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual page 148

5975 series
Table of Contents

Advertisement

4
General Maintenance
To remove the heater and sensor from the mass filter
Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
Procedure
1
2
3
4
Do not touch the mass filter contact leads. This could cause ESD damage to the side
CA UT IO N
board.
148
Large (8650-0030)
Small (8650-0029)
Vent the MSD.
Open the analyzer chamber.
Disconnect the mass filter heater and temperature sensor wires from the
feedthrough board.
Remove the mass filter heater assembly from the mass filter radiator.
5975 Series MSD Troubleshooting and Maintenance Manual