To clean the EI ion source
Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Solvents
• Ultrasonic bath
Preparation
1
2
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
5975 Series MSD Troubleshooting and Maintenance Manual
Large (8650-0030)
Small (8650-0029)
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Disassemble the ion source. See
Collect the following parts to be cleaned:
Repeller
Interface socket
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
General Maintenance
page
133.
(Figure
21)
4
135