General Maintenance

To Reinstall the Foreline Gauge

Materials needed
• Foreline gauge assembly (G1099-60545)
• Screwdriver, flat-blade, large (8730-0002)
Procedure
1
2
3
4
5
6
7
130
Connect a new foreline gauge assembly to the foreline hose.
Tighten the hose clamp.
Reconnect the foreline gauge cable to the foreline gauge.
Reconnect the foreline gauge assembly to the diffusion pump outlet.
Reconnect the MSD to the GC. See
If necessary, slide the foreline pump back under the analyzer chamber.
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
Pump down the MSD.
page 105
5975 Series MSD Troubleshooting and Maintenance Manual
and
page 106
for details.