To Drain The Foreline Pump - Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual

5975 series
Table of Contents

Advertisement

4
General Maintenance

To Drain the Foreline Pump

Standard foreline pump only
Materials needed
• Book or other solid object approximately 5 cm thick
• Container for catching old pump oil, 500 mL
• Gloves, oil- and solvent-resistant
• Screwdriver, flat-blade, large (8730-0002)
• Hex key, 5-mm (8710-1838)
Procedure
The foreline pump can cause burns if touched when operating. It has a safety shield
WARNING
to prevent the user from touching it.
1
2
The foreline pump may be hot.
WARNING
3
4
5
The old pump oil may contain toxic chemicals. Treat it as hazardous waste.
WARNING
98
Vent the MSD.
If necessary, slide the foreline pump to a safe, accessible location.
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
Place a book or other object under the pump motor to tilt it up slightly.
Remove the fill cap. See
Place a container under the drain plug.
Remove the drain plug. Allow the pump oil to drain out. The oil drains
faster if it is still warm.
Figure
7.
5975 Series MSD Troubleshooting and Maintenance Manual