Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual page 142

5975 series
Table of Contents

Advertisement

4
General Maintenance
To remove a filament
Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5-mm (8710-1570)
Procedure
1
2
3
4
The analyzer operates at high temperatures. Do not touch any part until you are sure
WARNING
it is cool.
142
Large (8650-0030)
Small (8650-0029)
Vent the MSD.
Open the analyzer chamber.
Remove the ion source. See the
Remove the filament(s) to be replaced
Operation
(Figure
5975 Series MSD Troubleshooting and Maintenance Manual
Manual.
23).