To disassemble the EI ion source
Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
Procedure
1
2
3
4
5
6
7
Video shows the standard Ion Source Assembly procedures.
NO T E
5975 Series MSD Troubleshooting and Maintenance Manual
•
Large (8650-0030)
•
Small (8650-0029)
Remove the ion source. See the
Remove the filaments
Separate the repeller assembly from the source body. The repeller
assembly includes the source heater assembly, repeller, and related parts.
Remove the repeller.
Unscrew the interface socket. A 10-mm open-end wrench fits the flats on
the interface socket.
Remove the setscrew for the lenses.
Push the lenses out of the source body.
Operation
(Figure
20).
General Maintenance
Manual.