9
Insert the flow cell capillaries and tighten them
fingertight. Use the wrench and torque adapter as
described on
Figure 76
fittings alternately.
11
Re-install the flow cell and connect the capillaries to the
union holder.
Agilent 1260 Infinity DAD and MWD User Manual
on page 222 and tighten the
Nano Flow Cell - Replacing or Cleaning
10
Reassemble the flow cell body to the holder.
Next Steps:
12
Perform a leak test with the flow cell outside of the
detector.
13
If no leak is observed, install the flow cell and you are
ready to work.
14
Make sure that the flow cell assembly is inserted
correctly and fits perfectly in the optical unit (especially
when PEEK capillaries are used).
10
Maintenance
221