9
Check for a centered holder vs. hole. If required adjust
with the holder screws.
Agilent 1260 Infinity DAD and MWD User Manual
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
Next Steps:
10
Reconnect the capillaries, see
Cell"
on page 200.
11
Perform a leak test.
12
Insert the flow cell.
13
Replace the front cover.
14
Perform a
page 183 or a
the correct positioning of the flow cell.
"Exchanging a Flow
"Wavelength Verification and Recalibration"
"Holmium Oxide Test"
10
Maintenance
on
on page 174 to check
217