Agilent Technologies 1260 User Manual page 122

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

6
How to optimize the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Table 29
Peak Width
[minutes]
<0.0025
>0.0025
>0.005
>0.01
>0.03
>0.05
>0.10
>0.20
>0.40
>0.85
122
Peak Width — Response Time — Data Rate
Response Time
[seconds]
0.025
0.05
0.1
0.2
0.5
1.0
2.0
4.0
8.0
16.0
Data Rate [Hz]
Detector
80
G1315C/G1365C only
80
G1315C/G1365C only
40
G1315C/G1365C only
20
G1315C/D and G1365C/D
10
G1315C/D and G1365C/D
5
G1315C/D and G1365C/D
2.5
G1315C/D and G1365C/D
1.25
G1315C/D and G1365C/D
0.62
G1315C/D and G1365C/D
0.31
G1315C/D and G1365C/D