9
Test Functions

ASTM Noise Test

ASTM Noise Test

The ASTM noise test determines the detector noise over a period of 20
minutes. The test is done with the flowcell removed, so the test results are not
influenced by solvent or pump effects. On completion of the test, the noise
result is displayed automatically.
This test is part of the detector self test only, see
N O T E

ASTM Noise Test Evaluation

ASTM Noise Test Evaluation
Limit is ± 0.02 mAU
Test Failed
Probable cause
1
2
178
Insufficient lamp warm-up time
Old or non-Agilent lamp.
"Self-test"
on page 162.
Suggested actions
Allow lamp to warm-up for at least 1 hour.
Exchange the lamp.