UV Ignition Failed

The UV lamp failed to ignite.
The processor monitors the UV lamp current during the ignition cycle. If the
lamp current does not rise above the lower limit within 2 – 5 seconds, the
error message is generated.
Probable cause
1
2
3
4
5
Agilent 1260 Infinity DAD and MWD User Manual
Lamp too hot. Hot gas discharge lamps may
not ignite as easily as cold lamps.
Lamp disconnected.
Defective UV lamp or non-Agilent lamp.
Defective detector main board.
Defective power supply.
Error Information
Detector Error Messages
Suggested actions
Switch off the lamp and allow it to cool down for
at least 15 minutes.
Ensure the lamp is connected.
Exchange the UV lamp.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.