Optimizing Selectivity; Quantifying Coeluting Peaks By Peak Suppression - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

Optimizing Selectivity

Quantifying Coeluting Peaks by Peak Suppression

In chromatography, two compounds may often elute together. A conventional
dual-signal detector can only detect and quantify both compounds
independently from each other if their spectra do not overlap. However, in
most cases this is highly unlikely.
With a dual-channel detector based on diode-array technology, quantifying
two compounds is possible even when both compounds absorb over the whole
wavelength range. The procedure is called peak suppression or signal
subtraction. As an example, the analysis of hydrochlorothiazide in the
presence of caffeine is described. If hydrochlorothiazide is analyzed in
biological samples, there is always a risk that caffeine is present which might
interfere chromatographically with hydrochlorothiazide. As the spectra in
Figure 58
where caffeine also shows significant absorbance. It would therefore be
impossible, with a conventional variable wavelength detector, to detect
hydrochlorothiazide quantitatively when caffeine is present.
Figure 58
Agilent 1260 Infinity DAD and MWD User Manual
on page 129 shows, hydrochlorothiazide is best detected at 222 nm,
Wavelength Selection for Peak Suppression
How to optimize the Detector

Optimizing Selectivity

6
129

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents