Setting the 8-bit Configuration Switch

Setting the 8-bit Configuration Switch
Setting the 8-bit Configuration Switch (with On-Board LAN)
The 8-bit configuration switch is located at the rear of the module. Switch
settings provide configuration parameters for LAN, serial communication
protocol and instrument specific initialization procedures.
All modules with on-board LAN, e.g. G1315/65C/D, G1314D/E, G4212A,
G4220A:
• Default is ALL switches DOWN (best settings) - Bootp mode for LAN.
• For specific LAN modes switches 3-8 must be set as required.
• For boot/test modes switches 1+2 must be UP plus required mode.
Figure 4
To perform any LAN configuration, SW1 and SW2 must be set to OFF. For details on the
N O T E
LAN settings/configuration refer to chapter "LAN Configuration".
24
Location of Configuration Switch
Agilent 1260 Infinity DAD and MWD User Manual