Automated Run Recovery In Case Of Temporary Communication Failures - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

5
Using the Detector
Special Settings of the Detector
Automated Run Recovery in case of temporary communication
failures
Table 24

Automated Run Recovery in case of temporary communication failures

Situation
Reaction
All OK
• Run in progress - Data Analysis
• Run / Rawdata
• Elapsed run time is running
• Spectra counter is running
• Data are stored on PC and on card
LAN breaks
• Run in progress - Data Analysis
• Run / Rawdata
• Error Power Fail
• Elapsed run time stops
• Spectra counter stops
• Data continues to be stored on card
LAN recovers
• Run in progress - Data Analysis
• Run / Rawdata
• Error Power Fail cleared
• Elapsed run time continues at actual time
• Spectra counter continues
• Data continues to be stored on PC and on card
• ChemStation tries already to add missing data (depends on the
data load).