Flow Connections To The Detector - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

Flow Connections to the Detector

Parts required
#
G1315-68755
Hardware required
Other modules
Preparations
Detector is installed in the LC system.
Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
The flow cell is shipped with a filling of isopropanol (also recommended when the
N O T E
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
The detector should be operated with the front cover in place to protect the flow cell area
N O T E
against strong drafts from the outside and to cover the deuterium lamp.
Some types of the Agilent deuterium lamps show a light ring during operation. This is not
harmful, refer to
Agilent 1260 Infinity DAD and MWD User Manual
Description
Accessory kit
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor and follow good
laboratory practice.
The amount of substances should be reduced to the minimal volume required for
the analysis.
Do not operate the instrument in an explosive atmosphere.
"UV-Radiation"

Flow Connections to the Detector

on page 272.
Installing the Module
3
53