Holmium Oxide Test Failed
Probable cause
1
2
3
4
5
Illegal Value From Temperature Sensor
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Probable cause
1
2
Agilent 1260 Infinity DAD and MWD User Manual
Lamps switched off.
Defective or dirty flow cell.
Defective filter assembly.
Defective achromat assembly.
Defective PDA/optical unit.
Dirty or defective sensor.
Detector is exposed to illegal ambient
conditions.
Error Information
Detector Error Messages
Suggested actions
Ensure the lamps are switched on.
Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
Exchange the filter assembly.
Exchange the achromat assembly.
Exchange the optical unit.
Suggested actions
Please contact your Agilent service
representative. Links:
Verify that the ambient conditions are within the
allowed range.