No value on temperature
display
Temperature in flask with
mobile Pt100 is too high
(> 65 °C)
Temperature in flask with
mobile Pt100 is too low
(< 0 °C)
No value on temperature
display
Temperature too high.
No error message
2013-02-19
Pt100 sensor or
sensor cable is
defect
Defect measuring
electronics
Defect measuring
electronics
Pt100 sensor or
sensor cable is
defect
Cooling is not
switched on
No power supply to
cooling system
Contact Infors representative
Contact Infors representative
Contact Infors representative
Contact Infors representative
Switch on cooling
Check fuse of cooling system,
replace if necessary
Connect unit to power supply
Multitron Pro - Incubation Shaker
Interferences
Qualified
technician
Qualified
technician
Qualified
technician
Qualified
technician
Operator
Operator
Operator
Page 161 of 185