Multitron Pro - Incubation Shaker
Operation
7.18 Operating the cooling system
Top-mounted cooling system 900 watts / 1200 watts
Side-mounted cooling system 380 watts
External cooling system
Page 132 of 185
Display show ON and changes to the current value of the
parameter, e.g. 65%.
NOTICE!
A setpoint of 70% rH is sufficient to prevent
evaporation.
Under ideal conditions, a setpoint of 85% rH can be
used.
NOTICE!
Formation of condensation indicates that the
maximum humidity has been reached. The setpoint
value must be reduced.
The cooling system has its own power supply. The cooling system
is switched on with the green push-button on the left-hand side of
the cooling module. The push-button lights up when the cooling
system is switched on.
The incubation shaker supplies power to this cooling system. The
cooling system is ready for operation when the incubation shaker is
switched on.
The external cooling system is supplied via a pressurised line from
the in-house cooling system. The external cooling system is ready
for operation when the incubation shaker is switched on.