Multitron Pro - Incubation Shaker
Operation
Page 118 of 185
2.
Press SET key.
Setpoint lights up.
3.
Press PLUS or MINUS key to set the desired value.
NOTICE!
The parameter values do not have to be confirmed.
They are accepted directly by the control system.
Setpoint is displayed in the upper display.
If no keys on the operating panel are pressed within 2 seconds, the
display alternates between the setpoint value und the operating
status OFF.
4.
Press ON/OFF key to switch parameter on.
Parameter sign lights up.
Select next desired parameter.
If no key is pressed within 2 seconds:
The upper display changes to the current value of the
parameter, e.g. 30.0 °C.
If this value is above the setpoint, „High" lights up.
If this value is below the setpoint, „Low" lights up.