Multitron Pro - Incubation Shaker
Operation
Unlocking the tray manually
Procedure
Fig. 87
Page 126 of 185
If there is a fault (power failure or mechanical problem) and the
door cannot be completely opened, the tray can also be removed
manually (by two people, if possible).
1.
Unload the tray.
2.
Open the door as far as possible.
3.
Person 1: Press both hooks of the locking mechanism
backwards.
4.
Person 2: Lift the tray out of the locking mechanism.
5.
Pull the tray over the front stop on the table.
2013-02-19