Multitron Pro - Incubation Shaker
Setup and function
4.9 Option «Sticky Stuff» adhesive matting
Fig. 33
4.10 Option Direct Steam Humidification
Page 68 of 185
The «Sticky Stuff» adhesive matting is a very sticky mat that lines
the tray. It can be detached from the tray and replaced.
The adhesive matting can also be used on perforated trays.
The adhesive matting provides secure purchase for cultivation
flasks with a large base.
The tray is labeled on two diagonally opposed corners with an
information sticker on how to remove the cultivation flasks.
These stickers must not be removed under any circumstances.
The incubator shaker is equipped with a Direct Steam
Humidification.
The Direct Steam Humidification is switched on or off via the
operating panel.
A humidity sensor, located on the rear wall of the incubation
chamber, measures the humidity.
Please note the following points:
The Direct Steam Humidification operates in one direction only.
It increases the humidity, but does not decrease it.
The Direct Steam Humidification only operates correctly at
temperatures up to 40 °C.
The temperature parameter must be switched on.
The cooling system switches off automatically when the Direct
Steam Humidification is switched on.