8.6 Illumination

8.6.1 Replacing fluorescent tubes

Procedure
Fig. 103
Fig. 104
2013-02-19
Replace defect fluorescent tubes as follows:
1.
Open the door.
2.
Switch off the unit.
3.
Rotate the tube by 90°.
4.
Pull the defect tube downwards out of both sockets.
5.
Push the new tube with the contact pins in a vertical position
upwards into the sockets.
6.
Rotate the tube by 90°.
Multitron Pro - Incubation Shaker
Maintenance
Page 147 of 185