Setting The Capillary Thermostat - Infors HT Multitron Pro Operating Manual

Incubation shaker
Table of Contents

Advertisement

6.7 Setting the capillary thermostat

Procedure
2013-02-19
Installation and initial operation
The capillary thermostat of the incubation chamber has a default
setting of 70 °C. This is 5 °C above the maximum operating
temperature of the unit, e.g. 65 °C.
The absolute tolerance of the thermostat is 2.5 °C.
1.
Set the thermostat to 70 °C.
2.
Shut the door of the unit.
3.
Switch on the unit.
4.
Press option key to select the temperature parameter.
The temperature parameter lights up.
5.
Press Set key.
Setpoint lights up.
6.
Press the PLUS key until the maximum operating temperature
(e.g. 50 °C) is set.
7.
Press the ON/OFF key to switch on the temperature control
system.
8.
Wait until the upper display has reached the setpoint value
(e.g. 50 °C).
Multitron Pro - Incubation Shaker
Page 91 of 185

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Multitron Pro and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF