Operating The Tray - Infors HT Multitron Pro Operating Manual

Incubation shaker
Table of Contents

Advertisement

CO
concentration is too high
2
Deviation > 1%
CO
value is fluctuating,
2
setpoint is not constant
Humidity is too low
Deviation > 10%
Humidity is too high
Deviation > 10%

7.11 Operating the tray

2013-02-19
CO
inlet pressure is
2
too high
Inlet pressure or flow
rate is too high
Water reservoir is
empty
Sterile filter (optional) is
blocked, e.g. after dry-
running
Temperature is too
high
Temperature is too low
Power failure
Unit was switched off
with activated
parameters
CAUTION!
Risk of damaging the door and the door
mechanism if the opened door is subjected to a
surface load of 26 kg or to a large localised
load.
A surface load of more than 26 kg or a large
localised load on the door can damage both the
door itself and the door mechanism.
Therefore:
Multitron Pro - Incubation Shaker
Reduce inlet pressure
Decrease inlet pressure or flow
rate
Fill water reservoir
Exchange sterile filter
Reduce the temperature setting
Increase the temperature setting
Open door to let wet air escape.
No actions required
No actions required
Do not place objects weighing more than 26 kg
on the door
Do not sit on the door
Do not use the door as a support
Operation
Operator
Operator
Operator
Operator
Operator
Operator
Operator
Page 123 of 185