8.8 Replacing the feed hose of the Direct Steam Humidification
Replacing the feed hose
Procedure
Fig. 107
8.9 Replacing the «Sticky Stuff» adhesive matting
Procedure
2013-02-19
Replace a damaged feed hose of the Direct Steam Humidification
as follows:
1.
Keeping the O-ring pressed down, pull the hose out of the
connector.
2.
Push the new hose into the connector (the O-ring does not
need to be pressed).
NOTICE!
Regular disinfection may reduce the adhesiveness
of the «Sticky Stuff» matting.
Replace the adhesive matting as follows:
1.
Release the adhesive matting on one side of the tray and then
pull upwards at an angle.
0
2.
Degrease the tray with acetone.
3.
Thoroughly moisten the tray with water.
4.
Wet the new matting and place on the tray (see the installation
instructions for «Sticky Stuff»).
Multitron Pro - Incubation Shaker
Maintenance
Page 149 of 185