Calibration For Level Measurement And Limit Detection; Function Group "Safety Settings" (*2) - Endress+Hauser Gammapilot M FMG60 Operating Instructions Manual

Radiometric measurement
Hide thumbs Also See for Gammapilot M FMG60:
Table of Contents

Advertisement

Commissioning
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
58
6.5

Function group "Safety settings" (*2)

group selection *2È
Âsafety settings
temp. compensation
linearization
6.5.1
"Output on alarm" (*20)
outp. on alarm
*20
MIN -10% 3.6mA
MAX 110% 22mA
Hold
This function determines what value the output of the Gammapilot M assumes in the event of an
alarm condition.
(*20)
MIN
MAX
Hold
User-specific
(can only be selected for HART
devices)
6.5.2
"Output on alarm" (*21)
outp. on alarm
*21
22.00 mA
This function is used to specify what user-specific value the current output should assume in the
event of an alarm condition. The value is entered in mA.
This function is only available for HART devices. It is only active if the "user-specific" option has
been selected in the "output on alarm" (*20) function.
Range of values
3.6 mA to 22 mA
Output on alarm
4 to 20 mA with HART
3.6 mA
22 mA
The last measured value is held.
As defined in "Output on alarm" (*21)
Gammapilot M FMG60 with HART
PROFIBUS PA
FOUNDATION Fieldbus
-99999
+99999
Not possible
Endress+Hauser

Advertisement

Table of Contents
loading

Table of Contents