System And Process Error Messages - Endress+Hauser PROline prosonic flow 93 Operating Instructions Manual

Profibus-dp/-pa ultrasonic flow measuring system
Hide thumbs Also See for PROline prosonic flow 93:
Table of Contents

Advertisement

PROline Prosonic Flow 93 PROFIBUS-DP/-PA
Device status and
Device status message
diagnosis message
(Display)
(Control system)
ROM / RAM failure
S CRITICAL FAILURE
$
Amplifier EEPROM
S AMP HW EEPROM
$
failure
Endress+Hauser
9.2

System and process error messages

General notes
The measuring device always assigns system and process errors which occur to two
types of error messages, resulting in different weightings.
Error message type "Fault message":
x If this message occurs, measuring operation is immediately interrupted or stopped!
x Display on the PROFIBUS o Fault messages are relayed to downstream function
blocks or higher-level process control systems with the status "BAD" of the
corresponding process variable.
x The fault messages can be displayed in the automation system by means of the
manufacturer-specific diagnosis message.
x Local display o A lightning symbol ($) flashes on the display.
Error message type "Notice message":
x Normal measuring operation continues despite this message!
x Display on the PROFIBUS o Notice messages are relayed to downstream function
blocks or higher-level process control systems with the status "UNCERTAIN" of the
corresponding process variable.
x The fault messages can be displayed in the automation system by means of the
manufacturer-specific diagnosis message.
x Local display o An exclamation mark (!) flashes on the display.
Serious system errors, e.g. electronic module defects, are always categorised and
displayed as "Fault messages" by the measuring device. On the other hand, the
measuring system interprets simulations and positive zero return as "Notice messages".
Error messages in the configuration program (Class 2 master) o see table
In Prosonic Flow 93, system/process errors are recognised and reported in the
Transducer and Analog Input Block. The following table contains a list of device status
messages pertaining to the Analog Input Blocks (PROFIBUS Profile 3.0) as well as a
description of the possible device status messages on the display.
Error messages on the local display o see table
No.
1
# 001
11
# 011
Initial status
Measured
Analog Input Block/
value Quality /
Totalizer Block
Substatus /
Alarm limit
device failure
BAD
0x0F
constant
device failure
BAD
0x0F
constant
9 Trouble-shooting
Cause/remedy
Cause:
System error. ROM / RAM error.
Error when accessing the program
memory (ROM) or random access
memory (RAM) of the processor.
Remedy:
Replace the amplifier board.
o
Spare parts
Page 136
Cause:
System error. Measuring amplifier
has faulty EEPROM
Remedy:
Replace the amplifier board.
o
Spare parts
Page 136
125