9 Troubleshooting
34
9
Troubleshooting
9.1
System error messages
Code
Error description
101 A
Checksum error
Total reset and recalib.
req.
102 A
Checksum error
Total reset and recalib.
req.
125 A
Sensor defective
W103
Initialisation active
Tab. 1
System error messages
For other error possiblilities, see KA ... (Basic device) table page 39.
Liquiphant M/S (HT), FEL 50 A PROFIBUS PA
Cause
• Device switched off before
data was saved
• EMC problem
• E2PROM defective
• Fork removed
• Chopping frequency reached
Device start-up after reset
Action
• Reset
• If alarm still present after
reset, replace electronics
• Reset
• Avoid EMC problems
• If alarm still present after
reset, replace electronics
• Insert fork
• Check fork for blockage and
remove same as necessary
Wait
Endress+Hauser