Diagnosis And Troubleshooting; System Error Messages - Endress+Hauser Smartec CLD132 Operating Instructions Manual

Measuring systems with inductive sensor for conductivity and concentration measurement in the food industry
Hide thumbs Also See for Smartec CLD132:
Table of Contents

Advertisement

Smartec CLD132/134 PROFIBUS PA/DP
NAMUR
Error
Description
class
no.
Failure
E001
Memory error
Failure
E002
Data error in EEPROM
Failure
E003
Invalid configuration
Failure
E007
Faulty transmitter
Failure
E008
Sensor or sensor
connection faulty
Failure
E010
Temperature sensor
defective
Failure
E025
Limit value for airset
offset exceeded
Failure
E036
Calibration range of
sensor exceeded
Failure
E037
Below calibration
range of sensor
Failure
E045
Calibration aborted
Failure
E049
Installation factor
exceeded
Failure
E050
Installation factor
undershot
Failure
E055
Measuring range of
the main parameter
undershot
Failure
E057
Measuring range of
the main parameter
exceeded
Failure
E059
Temperature range
undershot
Failure
E061
Temperature range
exceeded
Failure
E067
Limit switch set point
exceeded
Failure
E077
Temperature not
within the α-value
table
Failure
E078
Temperature not
within the
concentration table
Failure
E079
Conductivity not
within the
concentration table
Funct.
E101
Service function active
check
Endress+Hauser
9

Diagnosis and troubleshooting

9.1

System error messages

The DIAGNOSIS and DIAGNOSIS_EXTENSION parameters are generated from the device-
specific errors.
DIAGNOSIS
01 00 00 80 - DIA_HW_ELECTR
10 00 00 80 - DIA_MEM_CHKSUM
00 04 00 80 - DIA_CONF_INVAL
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
00 20 00 80 - DIA_MAINTENANCE
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
20 00 00 80 - DIA_MEASUREMENT
00 20 00 80 - DIA_MAINTENANCE
00 04 00 80 - DIA_CONF_INVAL
00 04 00 80 - DIA_CONF_INVAL
0 04 00 80 - DIA_CONF_INVAL
Diagnosis and troubleshooting
DIAGNOSIS_
Measured value status
EXTENSIO
Quality
01 00 00 00 00 00
BAD
02 00 00 00 00 00
BAD
04 00 00 00 00 00
BAD
08 00 00 00 00 00
BAD
10 00 00 00 00 00
BAD
20 00 00 00 00 00
BAD
40 00 00 00 00 00
BAD
80 00 00 00 00 00
BAD
00 01 00 00 00 00
BAD
00 02 00 00 00 00
BAD
00 04 00 00 00 00
BAD
00 08 00 00 00 00
BAD
00 10 00 00 00 00
UNCERT
AIN
00 20 00 00 00 00
UNCERT
AIN
00 40 00 00 00 00
UNCERT
AIN
00 80 00 00 00 00
UNCERT
AIN
00 00 00 04 00 00
UNCERT
AIN
00 00 01 00 00 00
BAD
00 00 02 00 00 00
BAD
00 00 04 00 00 00
BAD
-
1)
Sub-status
Hex
device failure
0C
device failure
0C
device failure
0C
device failure
0C
sensor failure
10
sensor failure
10
configuration
04
error
configuration
04
error
configuration
04
error
configuration
04
error
configuration
04
error
configuration
5C
error
sensor
50
conversion not
accurate
sensor
50
conversion not
accurate
sensor
50
conversion not
accurate
sensor
50
conversion not
accurate
non-specific
40
configuration
04
error
configuration
04
error
configuration
04
error
-
35